Hamamatsu C12666MA micro-spectrometer (340–780 nm, 256 pixels) read by a Raspberry Pi Pico through an MCP3301 13-bit ADC, on the JII "C12666MA Board v2". Includes a live GUI for chlorophyll fluorescence: it calculates relative fluorescence from the incident-light region on a selected top spectrometer and the fluorescence region on a selected bottom spectrometer, then plots the ratio over time. Multiple spectrometers are synchronized for recordings of minutes to days. A terminal recorder uploads the spectra to an openJII experiment (README_recorder.md).
firmware/main.py MicroPython firmware (copy to the Pico as main.py)
c12666ma/ Python package: driver, analysis, recorder, GUI, openJII upload and download
calibration/ pixel -> wavelength calibration and its laser measurements
tools/hw_check.py speed / reliability / noise check of a connected board
tests/ unit tests (no hardware needed)
notebooks/ K. Tolsma's measurement and PAM notebooks, and a Digilent WaveForms test
kicad/ PCB v2, two variants: Pico on the back / on the front
legacy/arduino/ earlier Arduino Uno sketches, incl. the MCP3301 bench test
legacy/python-v1/ the first MicroPython firmware (MAIN.py) and playground notebooks
pip install -r requirements.txt
python -m c12666ma # the GUI
python -m c12666ma --simulate 3 # try it with three simulated spectrometers
- Spectrometers plugged in are connected at start-up; later ones with Auto-detect & connect. Select one in the table to set its name, gain, integration time, averages and LED.
- Start. With one spectrometer, "As fast as possible" streams one spectrum per integration time (about 190 spectra/s at the 5.2 ms minimum). Otherwise all spectrometers take one spectrum together at every period (see below).
- Cover the sensors or switch the light off and press Take dark. The yield needs a dark spectrum taken with the same gain, integration time and averages, and the table shows when one no longer matches ("redo").
- Drag the blue (
incident_light) and red (fluorescence) regions, or type their limits. They apply to all spectrometers. Select the top (incident-light) and bottom (fluorescence) scanners in the controls below the live graph; both need matching dark spectra. - The 3D spectrum history tab shows the most recent spectra of the selected spectrometer as a surface over wavelength, time and counts. It keeps up to 5000 spectra per spectrometer in memory, can show the latest 2–500, and updates while it is open.
- Start recording, optionally with a duration. It runs until stopped or for days; a new file part starts every 24 h (adjustable).
run_C12666MA_3D_GUI.py starts the GUI from any folder:
python path\to\spectrometer-C12666MA\run_C12666MA_3D_GUI.py
All spectrometers take a spectrum together at every tick. The PC sends each
trigger up to 0.5 s ahead, with the start time converted to that device's
own clock. The clocks are measured every 30 s (offset and crystal drift), so
the devices should start within about a millisecond even when the PC
stalls. Spectra of the same tick have the same seq in every file.
- Minimum period: set by the slowest spectrometer, its averages × integration time plus about 12 ms (readout and margins). The GUI shows which one.
- Different integration times: the windows are centred on the same moment; a 10 ms spectrum is taken in the middle of a 1 s one.
- Unplugged or reset devices are reconnected automatically (by their unique ID) and the recording continues; gaps show up as missed ticks.
- Long recordings: the PC is kept awake while recording, and the recording stops cleanly (with a note) if less than 500 MB of disk is left.
The live GUI calculates bottom fluorescence / top incident_light from
dark-subtracted sums in the selected wavelength regions. It combines readings
only when both scanners have the same synchronized frame sequence number. A
region "has signal" when its sum is more than k × its noise (k = 5 by
default; the noise comes from that scanner's dark frames). Both scanners need
a matching dark spectrum. The per-scanner rel_fluo_yield values in recording
files remain calculated from each individual spectrometer.
| Top incident_light | Bottom fluorescence | result |
|---|---|---|
| signal | any | bottom fluorescence / top incident_light |
| no signal | signal | empty + warning fluorescence without incident light |
| no signal | no signal | 0 |
Frames with a pixel at saturation in either region are flagged (orange in
the plot, ;saturated in the file). The saturation level is about 3100
counts at high gain and about 1950 at low gain; both can be adjusted in the
GUI.
Each recording is a folder named <name>_<date>_<time>:
SIM-1_000001_part001.csv one series per spectrometer (<name>_<end of its ID>),
SIM-1_000001_part002.csv a new part every 24 h
SIM-2_000002_part001.csv
session.json devices, settings, calibrations, counts, why it stopped
events.log connections, reconnects, settings changes, ...
Open a part with pandas.read_csv(path, comment="#"). The header lines
(# ...) hold the settings, regions and calibration. The kind column
separates the row types:
wavelength_nm: the wavelength of every pixel columnpx0..px255dark/dark_std: the dark reference (written whenever one is taken, and repeated at the top of every part)spectrum: one row per spectrum, with the timestamp (start of integration),t_s(seconds since the recording started, the same time base for all spectrometers),seq(the tick number), settings, optionallyincident_light,fluorescence,rel_fluo_yield,yield_status, and the raw counts (not dark-subtracted).
python -m c12666ma.record records with all connected spectrometers from
the terminal, for hours or days, and sends the spectra to an experiment on
openJII as an upload table. Each unit keeps its own
gain, integration time and averages.
python -m c12666ma.record asks everything: sign-in, experiment, settings
python -m c12666ma.record login (re)enter the openJII API key
python -m c12666ma.record upload FOLDER finish the uploads of an earlier recording
python -m c12666ma.download download all tables of an experiment into one JSON file
The spectra are written locally first (the recording folder described above) and uploaded every 15 minutes, one row per spectrum with its timestamp and settings. Without network they wait on disk and go out later. README_recorder.md explains the whole procedure: sign-in, laptop settings for runs of several days, the uploaded columns, starting from Task Scheduler, downloading the data again and troubleshooting.
python -m c12666ma.record records with every C12666MA spectrometer plugged
into the laptop, one spectrum per unit at a fixed interval, for hours or days.
Each unit can have its own gain, integration time and averages. Every spectrum
is saved on the laptop first and then uploaded to an experiment on
openJII, with its timestamp and all the settings it was
taken with.
No GUI is needed: the script asks its questions in the terminal, and remembers the answers for the next run.
- Before you start
- First recording, step by step
- Leaving it running for days
- Stopping, and finishing the uploads
- What is saved
- Starting without questions
- Troubleshooting
-
Python 3.12 or 3.13, with the packages of this repository:
pip install -r requirements.txtRun every command below from the repository folder (the one with this file).
-
Firmware v2.1 on every Pico. Copy
firmware/main.pyto the Pico asmain.py(see the main README). The script refuses older firmware. -
A name and a calibration per unit. Units are recognised by their unique ID and shown by the name stored on the device (set it once in the GUI,
python -m c12666ma). A unit without its own wavelength calibration still records, with a warning: its wavelengths are those of another unit (see Wavelength calibration). -
An openJII account and an experiment you are a member of. To upload into an existing table you need contribute access. To create a new table, which the first upload does, you need manage access (experiment admin).
-
Close the GUI, Thonny or anything else that has the spectrometers' COM ports open. A port can only be used by one program at a time.
Plug in the spectrometers and run:
python -m c12666ma.record
The script goes through four steps. Press Enter to accept the value in
[brackets].
1. Sign in to openJII. The first time, the script opens Account > API keys on openJII in the browser. Create a personal key there, copy it and paste it into the terminal; it is not shown while you paste. The key is checked and stored in your Windows profile, so later runs skip this step.
Sign in to openJII (prod): create a personal API key at
https://openjii.org/en-US/platform/account/api-keys
Open that page in the browser? (y/n) [y]:
Paste the API key (hidden):
openJII (prod): signed in as John Doe
2. Choose the experiment and the table. Pick a number from the list of your active experiments, or choose the last entry and paste the experiment's link from the browser. The spectra go into an upload table of that experiment; keep the default name unless you want separate tables.
Active experiments you are a member of:
1. test-experiment (5db72a32-1c50-4d11-b12c-0152cc65b5e6)
2. another one (paste its id or URL)
Record into experiment: 1
Upload table name in the experiment [c12666ma_spectra]:
Experiment: test-experiment (https://openjii.org/en-US/platform/experiments/5db72a32-...)
Table: c12666ma_spectra (will be created)
3. Set each spectrometer. Every unit found on USB is listed, then asked for
its gain (high or low), integration time in milliseconds, number of
averages (1 to 16) and whether its LED is on. The units can have different
settings.
2 spectrometer(s):
Fluor COM42 id E664B012CF234E35 gain high integration 100 ms averages 1 LED off
Top COM7 id E663589863441E2E gain high integration 50 ms averages 2 LED off
Fluor on COM42 (id E664B012CF234E35)
gain (high/low) [high]:
integration time in ms [100.0]:
averages (1-16) [1]:
LED on during the recording? (y/n) [n]:
4. Timing and files.
| Question | What it means |
|---|---|
| Period between spectra in seconds | All units take one spectrum together at every period. The minimum shown depends on the slowest unit: its averages × integration time, plus about 12 ms. |
| Upload to openJII every | How often the new spectra are sent. 15 minutes is a good default; the minimum is 1 minute. |
| Local folder for the recording | Where the files are written. Each recording gets its own subfolder. Default: C12666MA data in your user folder. |
| Duration | For example 36h or 3d. 0 records until you press Ctrl-C. |
| Note for the file header | Free text saved with every spectrum, e.g. the plant or the chamber. |
| Take a dark spectrum first? | Optional. Cover the sensors when asked; 20 dark frames per unit are recorded and uploaded as dark rows. |
The recording then starts. Every minute a status line shows what has been measured and uploaded:
Recording to C:\Users\you\C12666MA data\test-experiment_2026-10-28_22-16-18
Period 60 s, upload every 15m, until Ctrl-C
22:17:18 · tick 7 · Fluor 6 · Top 6 · 12 spectra on disk · uploads 0 confirmed, 1 waiting
2026-10-28T22:17:20+01:00 upload: sent 1 batch(es), 12 rows, upload 7d8098da
2026-10-28T22:18:24+01:00 upload 7d8098da: 12 rows queryable on openJII
The spectra appear in the experiment's data on openJII about a minute after each upload. A new table is listed in the experiment a few minutes after its first upload. Until then the next uploads wait and retry every minute; the terminal says the table is not listed on the platform yet, which is normal.
The script keeps the laptop awake while it records, reconnects units that were unplugged or reset, and starts a new local file every 24 hours. Windows can still interrupt it, so before a long run:
- Connect the charger.
- Closing the lid: set Control Panel > Power Options > Choose what closing the lid does to Do nothing when plugged in, or leave the lid open.
- Windows Update: pause updates for the length of the run (Settings > Windows Update > Pause updates). An update restart stops the recording.
- USB power saving: in Power Options > Change plan settings > Change advanced power settings, set USB settings > USB selective suspend setting to Disabled.
- Don't select text in the terminal window. In the classic Windows console, a selection pauses the program until you press Esc.
- Disk space: the recording stops cleanly, with a note in the log, when less than 500 MB is left on the disk.
Network outages don't matter. The spectra keep being written to disk, the status line shows OFFLINE, keeping local, and everything that was missed is uploaded once the connection is back.
Two units at one spectrum per minute use about this much disk space:
| Per day | |
|---|---|
| Local CSV files | about 4 MB |
Upload batches, kept in uploads\ |
about 10 MB |
Press Ctrl-C once. The script sends the spectra not uploaded yet, waits up to a few minutes for openJII to confirm them, and then tells you whether anything is left:
Stopping (Ctrl-C)...
Uploading the last batch and waiting for the platform's confirmation...
Uploads: {'completed': 96}
Local files: C:\Users\you\C12666MA data\test-experiment_2026-10-28_22-16-18 (2880 spectra)
If the network was down, or the laptop crashed or restarted during a recording, send what is left later with the folder of that recording:
python -m c12666ma.record upload "C:\Users\you\C12666MA data\test-experiment_2026-10-28_22-16-18"
This also recovers the spectra written just before a crash. Running it on a recording that is already fully uploaded does nothing, so it is always safe.
A restarted laptop does not resume the recording by itself. Start a new recording; it goes into a new folder, and you finish the old folder's uploads with the command above.
Each recording folder holds:
Fluor_234E35_part001.csv spectra of one unit (<name>_<end of its ID>), a new part every 24 h
Top_441E2E_part001.csv
session.json units, settings, calibrations, number of spectra, why it stopped
events.log connections, reconnects, uploads, errors
uploads\ the batches sent to openJII and their status (ledger.json)
The CSV format is described in the main
README. Open a part with
pandas.read_csv(path, comment="#").
Each spectrum is one row in the experiment's upload table, by default
c12666ma_spectra. Every row describes itself completely:
| Column | Content |
|---|---|
kind |
spectrum, or dark for a dark reference |
timestamp_utc, timestamp_local, unix_time |
start of the integration |
t_s |
seconds since the recording started, the same time base for all units |
tick |
measurement number; spectra with the same tick were taken together |
session_id |
the recording folder name |
note |
the note entered at the start |
device_uid, device_name, port, firmware |
which unit took it |
integration_ms, integration_us, n_avg |
integration time and number of averages |
gain, high_gain |
high or low |
led |
whether the unit's LED was on |
flags, flag_names, adc_errors |
quality flags of the frame, e.g. late or led_changed |
clk_hz, n_pixels |
sensor clock and number of pixels (256) |
wavelength_coeffs, wavelength_nm |
the unit's calibration, and the wavelength of every pixel |
counts |
the 256 raw values, not dark-subtracted |
Dark rows hold counts_mean and counts_std instead of counts, plus the
number of frames averaged in n_frames.
Every question has a flag. With --yes the script asks nothing and uses the
flags, then the answers remembered from the last run. This is how to start it
from a batch file or from Windows Task Scheduler.
python -m c12666ma.record --yes --experiment "test-experiment" --table c12666ma_spectra ^
--device Fluor=high,100,1 --device Top=low,50,4,on --period 60 --upload-every 15m ^
--duration 5d --folder "D:\spectra" --note "chamber 2"
| Flag | Meaning |
|---|---|
--experiment |
experiment ID, link, or exact name |
--table |
upload table name (default c12666ma_spectra) |
--device NAME=gain,ms,averages[,on] |
settings of one unit, by name or ID; repeat for each unit |
--all gain,ms,averages[,on] |
the same settings for every unit |
--period |
seconds between spectra |
--upload-every |
e.g. 15m, 1h |
--folder, --prefix |
where to record, and the start of the folder name (default: the experiment name) |
--duration |
e.g. 36h, 3d; without it, until Ctrl-C |
--note |
note saved with every spectrum |
--dark [N] |
take N dark frames per unit first (default 20). With --yes it measures straight away, without asking you to cover the sensors. |
--part-hours |
hours per local file part (default 24) |
--no-upload |
record on the laptop only |
--env dev |
use the openJII development platform instead of openjii.org |
--yes needs a stored API key. Run python -m c12666ma.record login once
first.
Task Scheduler: create a task with the action Start a program. Set
Program to the full path of python.exe, Arguments to everything after
python in the command above, and Start in to the repository folder.
| Message | What to do |
|---|---|
no spectrometer found on USB |
Check the cables, and close the GUI or Thonny if they are open. |
firmware too old |
Copy firmware/main.py to that Pico. |
warning: Top has no calibration of its own |
The unit records with another unit's wavelengths. Calibrate it when possible. |
your account cannot contribute data |
Ask the experiment's admin to add you as a member. |
table ... does not exist ... (needs manage access) |
Ask an admin to make the first upload, or use a table name that already exists. |
no valid openJII API key |
No key is stored, or openJII no longer accepts it. Run python -m c12666ma.record login. |
the table is not listed on the platform yet |
Normal for a new table during its first minutes. The upload retries by itself. |
OFFLINE, keeping local in the status line |
No connection to openJII. Nothing to do: the uploads catch up later. |
LOST next to a unit in the status line |
The unit was unplugged or reset. It is reconnected automatically; the gap shows as missed ticks. |
upload ... FAILED on the platform |
openJII refused that batch and it is not resent. The spectra are still in the local CSV files and in uploads\. Report the upload ID shown. |
The stored key and answers are in %APPDATA%\c12666ma\openjii.json. To use
another key or account, run python -m c12666ma.record login. If the laptop
is shared or lost, delete the key on openJII under Account > API keys.
from c12666ma import Spectrometer
with Spectrometer.open() as spec: # finds the Pico automatically
spec.set_integration_us(20_000)
frame = spec.read_frame() # frame.counts: 256 values
spec.start_stream(period_us=0)
frame = spec.frames.get() # frames keep arriving until stop_stream()
spec.stop_stream()Copy firmware/main.py to the Pico as main.py, with Thonny or with
mpremote cp firmware/main.py :main.py. The serial protocol is documented
at the top of the file. Commands are text lines (set_integration,20000,
spec,raw, ...), and the v1 commands still work. Settings stay in RAM until
you send save. Synchronised acquisition needs firmware v2.1 (idn); the
GUI says so if a device has an older one.
The sensor clock (200 kHz), the start pulses and the ADC conversions are generated by the Pico's PIO state machines, and DMA moves the samples to RAM. The readout therefore doesn't depend on Python timing. Per the datasheet, the integration time is the interval between start pulses, so it is exactly N clocks for every pixel. Readout overlaps the next integration, so streaming has no dead time.
Measured on the board in October 2026 (python tools/hw_check.py, 60 s
run):
| frame rate at the 5.175 ms minimum integration | 193.2 spectra/s, period exact to 1 µs |
| dropped frames / checksum errors / ADC framing errors | 0 / 0 / 0 (11 593 frames) |
| device CPU load at that rate | 48 % |
| dark noise | ~11 counts rms per pixel (old firmware: same median, but outliers up to 350) |
The noise floor (~11 mV rms) is set by the analog front end, not the sensor (datasheet readout noise 0.5 mV).
There is no Hamamatsu inspection sheet for this unit, so
calibration/fit_wavelength.py fits a weighted quadratic to:
- a green (532.0 nm) and a red (650 nm) laser pointer (
calibration/*_laser.csv); - the chlorophyll fluorescence peaks (685 / ~737 nm) in 13 leaf recordings, which constrain the curvature and the red end.
The fit:
nm = 327.15 + 2.54396·p − 0.0024563·p² (p = pixel 0..255)
Expect about ±1 nm near 532 nm and a few nm up to 745 nm. Below 532 nm the curve is extrapolated, so a blue line would improve it: a violet (405 nm) laser, or a fluorescent lamp (Hg lines at 404.7 / 435.8 / 546.1 nm).
calibration/wavelength_calibration.json is the reference. When the GUI
connects, it writes these coefficients to the device's flash if the device
holds different ones. Scripts can read them back with get_wl_coeffs.
Every sensor needs its own calibration. A calibration file applies to the
unit whose ID is in its unit_uid field (the ID is in status and in the
GUI). Save a new unit's fit as calibration/wavelength_calibration_<name>.json
with its unit_uid. A unit without its own file is shown with another unit's
calibration, marked "approximate", and that calibration is not written to it.
- MCP3301 DOUT (5 V logic) is wired straight to GP16. The RP2040 is not 5 V tolerant. For the next revision, add a divider or a 3.3 V buffer; a ~2.2 kΩ series resistor is the quick fix.
- The video output reaches the ADC through 5 kΩ with no buffer amplifier (the datasheet recommends a CMOS op-amp buffer). This limits the ADC clock and probably adds noise.
- GP15 (the LED / excitation light) is not routed on the PCB and has to be wired separately.
- The schematic uses JII library symbols ("Library JII") that are embedded in the files but not included as a library.
python -m pytest # unit tests (simulated spectrometers, no hardware)
python tools/hw_check.py # with a board connected
python -m c12666ma --simulate 3 # the GUI with simulated spectrometers
python calibration/fit_wavelength.py # re-fit the calibration (the GUI copies it to the device)