new_Calibratron

The Calibratron rewritten for the ambit cmd-35 three-tier AS7341 chain (ambit/plans/AMBIT_COMMAND35_SPECPAR.md). Self-contained: it does not import the repo-root helpers.py.

spec_cal.py          pure - cmd-35 codecs, tier math, the tier-3 affine gate
quality.py           pure - the LED origin gate and the ADPD baseline gate
helpers.py           transport - discovery, Ambit link, references, record
run_calibratron.py   the bench sequence
tests/               78 tests, no hardware needed

Nothing here imports the repo-root helpers.py or calibration_quality.py.

Run: python run_calibratron.py · Test: python -m pytest tests -q

What it calibrates on the PAR chain

par_slope and par_intercept. Nothing else.

vector source this script
spec_offset[10] ams workbook — device-independent reads back, verifies
spec_sens[10] miniPar LR1-B seed (plan §7b) reads back, records
par_weight[10] miniPar Li-250A fleet seed, constrained fit (plan §7c) reads back, records
par_slope per-device intensity sweep fits, writes, verifies
par_intercept same sweep fits, writes, verifies

The three seeded vectors ship as firmware defaults. There is no code path here that writes them, deliberately rather than by omission: a per-device tier-2 fit from a single-lamp sweep scores R² −31 … −7443 with 434–779 % median error (plan §2). An intensity sweep moves the sample along one ray in 10-space, so it identifies a scale, not ten weights.

What was dropped from the old Calibratron

dropped why
cmd 31 / get_par / PAR — the whole legacy PAR path Integer Spec_COE weights on un-normalised counts at a pinned exposure, packed into a uint16 that wraps above ~11 % of full scale. A bright reading came back looking dark.
set_par_gain / set_spec writes spec_coef is frozen (plan decision 7). Still read and recorded — cmd 31 and deployed devices depend on it — never written.
ambit_spec_unscale, ambit_par_minipar_method, AMBIT_PAR_DISCREPANCIES All three existed to measure the gap between the legacy weighting and the MiniPAR's. Ambit now ships miniPar's fleet vector, so the gap they characterised is gone.
basic_count_divisor (seconds tick) Replaced by spec_cal.integration_time_ms. The two differ by 1000×; keeping both invites picking the wrong one.
Origin-forced gating of the PAR fit See below.
the repo-root calibration_quality.py Replaced by quality.py, which fixes four things that were wrong for this bench — see below. The root module stays for the legacy script.

Kept unchanged in substance: firmware flashing via firmware_fetch, device naming, actinic-LED gain, ADPD dark baseline, openJII publish.

The ADPD photodiodes during both sweeps

RECORD_ADPD_TRACES = True records an arrun2 trace — leaf, sun, s_630, r_630, env — at every point of both sweeps, and prints a per-point table. Nothing is fitted or written from these; they go in the record for later analysis.

The tier-3 sweep is the valuable one: it pairs the photodiodes with a calibrated PAR reference across the whole lamp range, which is a PAR response curve for leaf and sun obtained for free, at actinic = 0 so they see the same light the reference does. The LED sweep records them at actinic = setting, where the Ambit's own LED is the source.

Three things the implementation has to get right:

  • The ADPD pulse LEDs are zeroed before every trace. Without set_currents,0,0,0 the detector sees its own 620/720/IR pulses and sun / leaf stop being a measurement of the incident light. Re-asserted per trace rather than once per sweep: it costs one write, and if anything reset the device mid-sweep the currents would come back non-zero and every later reading would be contaminated with nothing in the record saying so.
  • The trace comes last at each point. It zeroes the pulse LEDs and drives the actinic, both of which change what the AS7341 sees, so it must not precede the cmd-35 read it is paired with.
  • Saturation is detected per channel. A pinned photodiode is the failure that quietly ruins a later analysis — the samples stay plausible and the spread collapses to zero. summarize_arrun flags any channel within 1 % of the 24-bit ceiling, the table marks it !, and assess_adpd_sweep names the drive it pinned at. It also reports a dead or reversed channel. All reported, never a gate: a flat photodiode must not discard an otherwise good tier-3 fit.

Cost is about 0.5 s of trace plus overhead per point, ~14 s across both sweeps. Cheaper than the old script's version, which opened and closed the port per point and so paid a device reset and boot each time. s_730 / r_730 stay empty because the run is type 2 (no IR reflect); sun and leaf are populated only because ambient sub-sampling is 1.

Why the PAR fit needs its own quality gate

calibration_quality.assess_origin_fit fits y = k·x and uses a free-intercept fit as a failure detector — an intercept above 5 % of full scale means the origin model was wrong, so it rejects. Tier 3's intercept is a deliverable, so that logic inverts: the thing being measured becomes grounds for rejection.

The deeper cost is what origin-forcing does to the numbers. With a true b = 7.983 — the value plan §7c measured on the superseded OLS seed, 2.40 µmol of dark-offset term plus 5.61 µmol of miniPar's discarded tier-2 intercept — forcing through the origin folds the offset into the slope. At the bright end that is a sub-1 % error; at 120 tier-2 counts it is over 5 %, and it keeps growing as the light drops. A constant additive error becomes a proportional one, worst exactly where canopy and shade measurements live. tests/test_spec_cal.py::test_origin_forcing_would_misplace_a_real_intercept pins that.

Two further incompatibilities: R² scored against the origin-forced prediction gates on "did b happen to be small" rather than on linearity; and the non-negativity check rejects a legitimately slightly-negative par_tier2 at the dark point, which the seeded par_weight (negative on NIR) can produce. That is the one point the intercept is fitted from.

So spec_cal.assess_affine_fit fits both parameters, bounds-checks them against the firmware predicates (par_slope ∈ (0,100], |par_intercept| ≤ 500), scores R²/NRMSE against the affine prediction, permits negative inputs, and demotes the intercept-size question to a note keyed on the dimmest lit point rather than a rejection.

What quality.py fixes in the root calibration_quality.py

The root module's arithmetic is fine and the legacy script still uses it. Four things were wrong for the new approach, and quality.py is a straight replacement rather than a wrapper because two of them are API shape:

  1. Its low coefficient bound was inclusive where the firmware's is exclusive. valid_actinic_coefficient is > 0.01 && <= 1.0, and ambit's own test asserts "actinic lower bound is exclusive". The host gate was coefficient_min <= coefficient, so a fit of exactly 0.01 passed host QC and was then refused by the device — the bench would report a successful write with nothing changed. Same class as sending an unrounded par_slope.
  2. One gate was dead at the only call site. The old signature was (x, y, stimulus) and calibrate_led passed the LED settings as both y and stimulus, so "reference readings are not monotonic with the applied stimulus" compared the settings to themselves and could never fire. assess_led_fit(measured, settings) takes two arguments, which makes the mistake impossible instead of merely unlikely.
  3. A slightly negative dark reading failed the whole sweep closed. Any negative value tripped "calibration values must be non-negative". The MiniPAR reports par_raw * slope + intercept; with a negative intercept a genuinely dark reading comes back a shade under zero. That is noise. Small negatives are now clamped with a note; large ones still fail.
  4. Its thresholds shared names with different values in spec_cal. Both defined MIN_R2, MAX_NRMSE and MAX_MONOTONIC_REVERSAL, at 0.99/0.05/0.02 against 0.995/0.03/0.02. Two places to get a tuning edit wrong. Every threshold here is prefixed LED_* or ADPD_*, and a test asserts the names cannot collide.

The ADPD baseline gate moved out of the runner into the same file, so every numerical gate the bench applies now sits next to the firmware bound it mirrors.

The origin fit itself is kept, and is correct here: zero drive is zero light with no offset to absorb, so a free intercept is evidence against the model rather than a quantity to fit — the opposite of tier 3.

One honest caveat on what the LED step buys you: as of ambit@fix/spec_overflow, actinic_coef is validated, persisted and printed but never applied — AS_LED_Current() takes the requested setting directly. So the step characterises the LED rather than changing device behaviour. Worth running, worth not over-trusting. Noted in the function's docstring too.

The MiniPAR as reference instead of a Li-250A

Accepted. The cost is smaller than it looks, for a structural reason.

Ambit's seeded par_weight is miniPar's fleet vector, so at any single calibration spectrum both instruments compute w · s over the same weights. Whatever spectral bias w carries at that spectrum is common-mode and divides out of the ratio, leaving the fitted slope as (near enough) the pure optical-throughput ratio of the two stacks. The MiniPAR behaves as a transfer standard, and the fit's linearity is untouched — which is why the QC gates still mean what they say.

What it costs is a pure multiplicative bias: the reference MiniPAR's own PAR error at the calibration lamp's spectrum.

term figure source
MiniPAR in-domain 1.96 % median, R² 0.9983 plan §7c, 462 samples, 3.7–2033.7 µmol
MiniPAR cross-source 4.37 % median, 9.82 % worst device plan §2, LED-calibrated → scored on daylight
Li-250A absolute accuracy ~5 % Li-Cor spec, unavoidable either way
combined ~7 % typical, ~10 % worst in quadrature

The cross-source row is the honest one: the bench lamp is halogen, which is not in miniPar's daylight-dominated calibration set. So the choice takes you from ~5 % to ~7 %.

It is recoverable without redoing bench work. The record stores par_tier2 per sweep point, so one later Li-250A comparison on a single device yields a scale correction applicable to every stored sweep. That is why the payload carries the whole sweep and not just the fitted pair.

Two caveats the record captures rather than hides:

  • The sweep is not one spectrum. A halogen lamp from 0.4 A to 6.6 A shifts colour temperature by hundreds of kelvin. Both instruments see it and use the same weights, so it largely cancels — but spec_cal.spectral_drift measures how far it moved, because that bounds how far the fitted slope travels from this lamp.
  • par_slope far from 1.0 is informative. par_tier2 is already in µmol (plan §7c fitted w to produce them directly, and a tier-3 refit on miniPar data returns a = 1.0000). A slope outside 0.2–5.0 means ambit's optics differ from miniPar's by more than that, or the reference is misconfigured. Noted, not rejected — nobody has measured ambit optics yet.

A clipped point cannot be fitted — including the dark one

Found while checking the plan against the sweep design; now written into plan §7c.

Plan §7c justifies reusing miniPar's par_weight at all by showing that ambit's extra offset subtraction costs a constant Σ wᵢ·offsetᵢ — 1.09 µmol on the current seed, 2.40 on the superseded OLS one — which tier 3's intercept absorbs exactly. That constancy holds only while no channel is clipped. Once s saturates at zero the offset term stops being constant and the relation stops being affine — and the clip is the chain's only nonlinearity (§5).

This became live rather than theoretical when the real ams offsets replaced zeros: with a zero offset vector s == x always and nothing can ever clip. The offsets are only 0.28 – 2.02 raw counts at the pinned 2× / 139 ms, so the clip bites solely in a light-tight fixture — but there it bites on all ten channels at once, which is exactly the dark point of a sweep.

Concretely: a fully clipped dark reading has par_tier2 identically 0 against a reference of ~0, so it lies on the line only if par_intercept is 0. Folding it into the fit drags the intercept from 7.983 to 4.78 — 40 % low — while R² stays above 0.99, so no quality gate catches it. (tests/…::test_fitting_a_clipped_dark_point_would_bias_the_intercept.)

So spec_cal.usable_for_fit rejects any clipped point at any lamp drive, and the sweep gained a 0.4 A step: 0.0 A stays as the cheapest light-tightness check, and 0.4 A is a genuine low anchor inside the linear regime. If the dark point comes back unclipped it is fitted too — the rule keys on clip_mask, not on the drive.

The par_weight seed has a generation, and it changed

SEED_PAR_WEIGHT is the constrained miniPar fleet fit — lsq_linear on basic counts with F1–F8 ≥ 0 and Clear/NIR free, shrunk toward W_TARGET at λ=0.03, exported by miniPar/new_calibration_miniPAR/par_coeffs_fleet.json. SEED_GENERATION is minipar-2026-08-18-constrained.

It replaced a plain OLS fit on the same 462 samples. That one predicted in-domain PAR just as well (in-sample R² 0.9984 against 0.9982) but its signs were not physical — negative on F3, F5 and F8, positive on Clear — because the design matrix has condition number ≈ 451 on a daylight-dominated set. The practical cost is leverage on spectral shape rather than in-domain error: L1 norm 1172 against 457 for a 1.8× larger net response, and off-daylight it shows up directly — canopy PAR spread 6.2 % against 0.4 %, coefficient direction spread p95 62° against 8°.

Two things follow for this script:

  • A tier-3 fit is only interpretable against the tier-2 vector it sat on. The two vectors give tier-2 values within 5–8 % of each other on the bench lamp, so a re-swept par_slope moves by about that much and no gate would notice. Every record therefore carries seed_generation and par_weight_generation.
  • A device on older firmware must be named, not called unknown. A False in seed_match means "someone wrote a fitted vector here" everywhere else, so SUPERSEDED_PAR_WEIGHTS keeps the previous vector and SpecCal.par_weight_generation() reports minipar-2026-08-17-ols for it. read_par_provisional adds a reason saying to reflash before trusting a fit taken on it.

spec_sens and spec_offset were already bit-for-bit the current spectral_coeffs_fleet.json export and did not change. bit8 stays clear: this is still a fit on miniPar optics, not an ambit one. And the export still carries prior_is_placeholder: true, so expect one more generation once the CM weights are computed.

How this host reads the flags word

The firmware zones flags into two bytes of opposite polarity, and it is worth understanding before writing any host code against it:

low byte conditions, 1 = needs attention bit0 full scale · bit1 dark-offset clip · bit2 ASAT · bit3 acquisition fault
high byte calibration, 1 = confirmed bit8 par_weight is an ambit fleet fit · bit9 tier 3 stored for this device

An all-zero word — what a truncated or zeroed frame produces — therefore reads as "no fault reported, nothing confirmed calibrated", which is the pessimistic reading. Confirmation requires bits to be set; treating an unset bit as trustworthy is the failure the zoning prevents. So par_provisional here is not (bit8 and bit9), and spec_cal.py never collapses the two.

Two consequences for this script:

  • bit8 stays clear after a successful calibration, because AMBIT_PAR_WEIGHT_IS_AMBIT_FIT is compile-time false until the ambit Li-250A campaign lands. Every device the Calibratron produces will report provisional PAR. That is correct, and it is never a success criterion here — the criteria are the cmd 33/4 read-back and the confirmation pass.
  • bit9 is authoritative for "has this device been swept", keyed on NVS key presence. read_par_provisional still cross-checks it against the read-back vectors, but the check is directional: bit9 latches for the life of the NVS partition (there is no verb to delete the key — firmware plan §5a), so bit9 set over identity vectors is a legal state (a bench write, or a slope restored to exactly 1.0) that earns a printed reason and nothing more. Only the impossible direction — bit9 clear over non-identity vectors — aborts the sweep, because a fit taken on top of an unknown tier-2 state is not a calibration.

The reboot dump does not contain the PAR chain. The five vectors live outside ambit_calibration_info_t by design (decision 2), so the boot banner shows a device that looks fully calibrated while its PAR chain is untouched. Every step here reads cmd 33/4 alongside the dump, and read-back verification goes through cmd 33/4, never through ambit_reboot.

Writes have three outcomes, not two. report_spec_save answers "<what> saved and verified", "<what> rejected", or "<what> save failed: <ESP_ERR_…>". The third contains no negative keyword, so a host testing for "rejected" reads an NVS failure as a success. Acceptance is tested positively on saved and verified — one test that covers all three outcomes and any future rewording.

get_spec_cal is five labelled lines, not one CSV row: spec_offset: / spec_sens: / par_weight: (ten %.9g each) then par_slope: and par_intercept:. parse_spec_cal_text keys on the labels, never on position — a one-slot shift in these vectors is undetectable in the numbers, since every element is a plausible magnitude for its neighbour. %.9g round-trips an IEEE-754 float exactly, so the text mirror and cmd 33/4 agree bit for bit.

Bench sequence

probe cmd 35 (80 B, format 1)          -> abort tier 3 if absent; no legacy fallback
read cmd 33/4                          -> previous tier 3, for rollback
verify firmware math from raw[]        -> catches tick / gain ordinal / bank / NIR-Clear
sweep 0.0 .. 6.6 A                     -> par_tier2 vs reference PAR, per-point masks
affine fit + QC
write par_slope, par_icept             -> verify via cmd 33/4, roll back on mismatch
confirm at 0.8 / 3.0 / 6.6 A           -> device par vs reference, closed loop

The firmware-math check is the one that earns cmd 35's design: raw[] stays on the wire so the host can reproduce every derived field, and a single comparison covers all four footguns in plan §8 — each of which is otherwise invisible at the pinned exposure. If it fails, the run stops before the fit, because a fit taken on wrong arithmetic would encode the bug as a calibration coefficient.

Every MiniPAR read is a coin flip unless the port is opened cold

The MiniPARs are ESP32-class boards with reset wired to DTR/RTS, so pyserial's default open — which asserts both — reboots the instrument. The command then goes into a booting device and the reply is one of:

Saved PC:0x40053b60          bootloader output
load:0x3fcd5810,len:0x438    bootloader output
error:unknown_command        the command line, chewed up by the reset
17.056                       the previous question's answer, still buffered

Only the last one is dangerous: it parses. A stale par reading pairs the wrong reference with the Ambit's par_tier2, and the fit absorbs it silently.

So helpers._query opens cold (open_serial_no_reset, DTR/RTS set before open), drops whatever is already buffered, skips bootloader lines without spending an attempt, and re-asks up to QUERY_ATTEMPTS times whenever the reply is an error: or the wrong shape for the question — every getter passes what shape it expects. _command is deliberately left on the default open: its only user is the Kiprim DC source, which has no reset on those pins and may hold its output until DTR is asserted.

A reference that still will not answer raises ReferenceUnavailable rather than returning a number: the tier-3 sweep aborts with status: reference_unavailable, writes nothing, and drops the lamp to 0 A in a finally. A partial sweep against a silent reference is not a calibration.

BAD COMMAND: one space in a command template

set_currents is implemented in every firmware this bench flashes (ambit/src/do_command.h, case hash("set_currents"), printing Currents set to %d, %d, %d). Six BAD COMMAND replies to it per run were not a missing verb. They came from the trailing space in arrun1 / arrun2's padding, ", , ".

The device reads each command with Serial_Input_Chars(choose, ":,", 200, ...), and serial.cpp gives that reader three relevant properties:

  • : and , end a token and are discarded;
  • CR/LF are skipped without being stored and without restarting the timer;
  • anything else is stored, and storing restarts a 200 ms inter-character window.

A space is printable, so it is stored as token byte 0 and opens that window. Whatever the host sends inside it joins the same token, so set_currents,0,0,0, arrives as " set_currents" - and do_command() drops any token whose first character fails isalnum(), silently. The verb is gone. Its three 0 arguments are then read as commands, and a digit-leading token goes through atoi into a switch with no numeric cases: BAD COMMAND, three times.

Why the first trace of each sweep always worked: set_actinic sleeps 0.3 s and flushes, which outlasts the 200 ms window. arrun returned the instant it saw Data sent and did not - so traces 2..7 lost their set_currents and the ADPD pulse LEDs stayed live through the rest of the sweep.

Fixed at the root (padding is now ", , ", which stores nothing) and defended at the host (AmbitLink._settle_console, CONSOLE_TOKEN_SETTLE_S = 0.25, after every trace). AmbitLink.text also takes expect= now, which was a real fix for stale-line attribution but was NOT this bug - there was no acknowledgement to find, because the verb never reached the parser.

tests/test_console_tokenizer.py models the reader's three rules and asserts the next verb survives each padded template; it also asserts the old padding fails, so the guard is shown to catch what it was written for.

The consequence was not cosmetic: pulse_currents_zeroed was False, so leaf and sun in those traces measured the Ambit's own pulse LEDs alongside the incident light. That flag now travels beside adpd_stats (raw samples are dropped, provenance is not), the station block reports pulse_leds_zeroed_confirmed rather than asserting the intent, and one line under the table tells the operator the numbers are contaminated.

leaf and sun face different directions

Which photodiode responds is fixture geometry, not detector health:

sweep leaf sun
halogen lamp, 0.0-6.6 A span 69, wanders span 927, monotonic
Ambit actinic LED, 10-250 790 -> 1134, monotonic flat at ~655

So assess_adpd_sweep takes responders= and only the channel aimed at the source gets the flat / non-monotonic notes - ("sun",) for the lamp sweep, ("leaf",) for the LED sweep. The other channel is still recorded, and saturation is still reported for both: a pinned photodiode is a fault wherever it points. Before this, every run reported the geometry as a fault, which is how a genuinely flat channel would have gone unnoticed.

env is not a photodiode at all - it is the MLX object temperature, sampled once per trace rather than once per pulse. Its +-0.0 is arithmetic on a single value, not an unusually quiet channel, and the object cannot warm measurably inside a 0.5 s trace.

A missing instrument stops the bench

REQUIRE_ALL_DEVICES (default on). A role that does not answer used to degrade quietly: tier 3 printed one [skip] line and the run still saved and uploaded a record. Discovery failing is a bench fault, so it now aborts before the reboot dump and before the rename prompt - nothing has been touched at that point.

Discovery also says why now. A port that is present but will not open used to be a logger.debug line, which made it indistinguishable from an unplugged instrument; both are warnings, and the "no device for role" message lists every port that was actually asked.

Status

The firmware side this host talks to is built, on ambit@fix/spec_overflow: the cmd-35 payload, cmd 33/4, the five text setters, get_spec_cal, the predicates, the two-zone flags word, and the seeded spec_sens / par_weight in nvs1.h are all in place. Verified against the source, not assumed.

What is still open is measurement, not code (plan §11): no vector here is fitted on ambit hardware. AMBIT_PAR_WEIGHT_IS_AMBIT_FIT is false, so bit8 stays clear and every device this bench produces reports provisional PAR — as it should, until an ambit Li-250A campaign replaces §7c.

Desktop GUI

calibratron_gui.py is a tkinter front-end over run_calibratron.py: the same per-device pass (flash, name, tier-3 PAR, actinic LED, record, publish), with the bench instruments confirmed from the cached port roles for every device and the openJII sign-in kept for the whole session.

Download the ZIP for your platform from this repository's GitHub Releases:

  • calibratron-windows-x64.zip: extract the whole folder and run calibratron.exe.
  • calibratron-linux-x64.zip: extract and run calibratron/calibratron (Ubuntu 22.04 or newer).
  • calibratron-macos-arm64.zip: extract and open calibratron.app (Apple Silicon).

Keep the bundle's supporting files beside the executable. Python and esptool do not need to be installed separately; the serial drivers and the bench hardware are still required. The bundles are unsigned, so Windows and macOS may show an unknown-publisher warning. Intel macOS is not packaged.

For a source checkout, install requirements-gui.txt (Linux also needs the distribution's python3-tk), then run python calibratron_gui.py. Press Rescan bench, select the steps, and press Start calibration.

Packaged applications keep calibrations/, firmware_cache/ and the .port_roles.json port hints beneath a per-user data directory (runtime_paths.data_dir()):

Platform Data directory
Windows %LOCALAPPDATA%/Calibratron
macOS ~/Library/Application Support/Calibratron
Linux ${XDG_DATA_HOME:-~/.local/share}/Calibratron

Set CALIBRATRON_DATA_DIR to override it. A source checkout uses the checkout folder itself. The firmware release is resolved once per session: the first device asks GitHub, every later device reuses that answer. The downloaded firmware cache is kept when the GUI closes, so the next session re-proves the release online and falls back to the cached copy if GitHub will not answer (anonymous API calls are limited to 60 per hour per public IP). API-key settings live in the separate per-user configuration directory managed by openjii_auth.py.

GUI release pipeline

.github/workflows/gui-release.yml tests and builds Windows x64, Linux x64 and macOS ARM64 bundles on pull requests, pushes to main, and manual runs. Every platform runs the packaged executable's --smoke-test, which constructs the GUI without signing in, loads the bundled CA roots, and exercises the bundled esptool without touching a serial port. PR and manual runs leave the ZIPs as Actions artifacts.

After all three builds pass on main, the same run's ZIPs and SHA256SUMS are uploaded to a draft release and then published. Tags are calibratron-vX.Y.Z and the patch number increments automatically; re-running an already published commit does not create another release. Publication uses the built-in GITHUB_TOKEN: no extra secret and no hand-made tag. The workflow builds the merged main commit, so merge commits, squash merges and direct pushes all release the same way.

Local rehearsal of what CI does:

python -m pip install -r requirements-build.txt
python -m pytest tests -q
pyinstaller --noconfirm --clean --onedir --windowed --name calibratron \
  --paths . --collect-all esptool packaging/launcher.py
dist/calibratron/calibratron --smoke-test     # Linux: xvfb-run -a ...

Automated packaging checks do not replace a physical bench test.